Calibration of Spectrophotometer


1.0   PURPOSE:   To ensure smooth functioning of Spectrophotometer.

2.0   SCOPE   :   The SOP applies to the calibration of Spectrophotometer of Shimadzu
                            Company model No. UV-1201

3.0   DEFINITION:  Nil.

4.0  RESPONSIBILITY :  QC officer.

5.0  PROCEDURE:

5.1  Switch on the machine & wait for auto calibration .
5.2  After auto calibration check the absorbance of potassium Dichromate Solution
VS at the wavelengths indicated in the following table which gives for each
Wavelength the exact value of A (1% 1cm) & the permitted limits. The tolerance of absorbance is ± 0.01

          Wavelength (nm)                     A(1% 1cm)                     Maximum Tolerance   
                 235                        -                124.5           -                  122.9 – 126.2
                 257                        -                 144.0           -                  142.4 – 145.2
                 315                        -                 48.6            -                    47.0 – 50.3
                 350                        -                 106.6           -                   104.9 – 108.2