1.0
PURPOSE: To
ensure smooth functioning of HPLC Machine.
2.0
SCOPE : The SOP applies to the calibration of
UV-VIS detector of Shimadzu
Company model No. SPD-10A vp.
3.0
DEFINITION: Nil.
4.0 RESPONSBILITY:
Quality Control officer.
5.0 PROCEDURES:
5.1
Press & hold down the “shift” key & turn the
power switch on.
5.2
Press the “func” key until lamp is displayed.
5.3
Press “1”,
then enter.
5.4
Press “func” until “Wave check” is displayed.
5.5
Press “enter”. The wavelength accuracy check is run
automatically.
CHECK GOOD should appear on the
screen when the test ends.
|