1.0 PURPOSE: To ensure smooth functioning of
Spectrophotometer.
2.0 SCOPE :
The SOP applies to the calibration of Spectrophotometer of Shimadzu
Company model No. UV-1201
3.0 DEFINITION:
Nil.
4.0 RESPONSIBILITY
: QC officer.
5.0 PROCEDURE:
5.1 Switch
on the machine & wait for auto calibration .
5.2 After
auto calibration check the absorbance of potassium Dichromate Solution
VS at the wavelengths indicated
in the following table which gives for each
Wavelength the exact value of A
(1% 1cm) & the permitted limits. The tolerance of absorbance is ± 0.01
Wavelength (nm) A(1% 1cm) Maximum Tolerance
235
- 124.5 - 122.9 – 126.2
257
- 144.0 - 142.4 – 145.2
315
- 48.6 - 47.0 – 50.3
350 - 106.6 - 104.9 – 108.2
|
Showing posts with label Spectrophotometer. Show all posts
Showing posts with label Spectrophotometer. Show all posts
Friday, May 17, 2013
Calibration of Spectrophotometer